Table of Contents

Color measurement with fixed or variable aperture
color2view
Color measurement with fixed aperture
- Objective evaluation of Color, 20° / 60° Gloss and lightfastness by combining 3 measurement technologies in one instrument
- Fixed aperture (32 mm) with extra-large measuring spot (22 mm) guarantees highly reliable readings
Technical Parameters
Download Documents
Category | Specification |
Measuring Capability | Color, Gloss, Fluorescence |
Color Measurement | |
Color Geometry | 45°c:0° |
Sample Port | 32 mm |
Measuring Area | 22 mm |
Spectral Range (Color) | 400–700 nm, 10 nm resolution |
Spectral Range (Fluorescence) | 300–760 nm, 10 nm resolution |
Measurement Range | 0–170% remission |
Repeatability (Color) | 0.01 ΔE94 (10 readings on white) |
Reproducibility (Color) | 0.1 ΔE94 (average of 12 BCRA II tiles) |
Color Systems | CIELab/Ch, Lab(h), XYZ, Yxy |
Color Indices | YIE313, YID1925, WIE313, CIE, Berger, Color Strength,Opacity, Metamerism, Grayscale |
Illuminants | A, C, D50, D55, D65, D75, F2, F6, F7, F8, F10, F11,UL30, CIE 015:2018 LED Normlicht |
Observer Angles | 2°, 10° |
Gloss Measurement | |
Measurement Range | 20° Geometry: 0 – 100 GU; 60° Geometry: 0 – 100 GU |
Aperture Size | 20° Geometry: 9 × 10 mm; 60° Geometry: 9 × 16 mm |
Repeatability | 20° Geometry: ± 0.2 GU (0–100 GU); 60° Geometry: ± 0.1 GU (0–20 GU), ± 0.2 GU (20–100 GU) |
Reproducibility | 20° Geometry: ± 1.0 GU (0–100 GU); 60° Geometry: ± 0.5 GU (0–20 GU), ± 1.0 GU (20–100 GU) |
Available for Aperture | 20° Geometry: 32 mm, 12 mm; 60° Geometry: 32 mm |

Color & Jetness measurement: Fixed or variable aperture
color2view Pro X
Color & Jetness measurement with variable apertures
- Objective evaluation of Color, 20° / 60° Gloss and lightfastness by combining 3 measurement technologies in one instrument
- Exchangeable apertures in 3 sizes – 32/22 mm, 12/08 mm, 05/04 mm – ensure maximum flexibility to measure large to very small samples
- Quality control of deepest blacks with an unsurpassed accuracy offering a special Jetness Mode
Technical Parameters
Download Documents
Category | Specification |
Measuring Capability | Color, Gloss, Fluorescence (32 mm, 12 mm), Jetness (32 mm) |
Color Measurement | |
Color Geometry | 45°c:0° |
Sample Port | 32 mm, 12 mm, 04 mm |
Measuring Area | 22 mm, 08 mm, 04 mm |
Spectral Range (Color) | 400–700 nm, 10 nm resolution |
Spectral Range (Fluorescence) | 300–760 nm, 10 nm resolution |
Measurement Range | 0–170% remission |
Repeatability (Color) | 0.01 ΔE94 (10 readings on white) |
Reproducibility (Color) | 0.1 ΔE94 (avg. of 12 BCRA II tiles) |
Color Systems | CIELab/Ch, Lab(h), XYZ, Yxy |
Color Indices | YIE313, YID1925, WIE313, CIE, Berger, Color Strength,Opacity, Metamerism, Grayscale |
Illuminants | A, C, D50, D55, D65, D75, F2, F6, F7, F8, F10, F11, UL30, CIE 015:2018 LED Normlicht |
Observer Angles | 2°, 10° |
Jetness Measurement | |
Jetness Indices | MY, Mc, dM, GY, Gc, dG |
Repeatability Jetness | Y ≤ ± 0.0005 |
Reproducibility Jetness | Y ≤ ± 0.003 (Y < 0.5) |
Gloss Measurement | |
Measurement Range | 20° Geometry: 0 – 100 GU; 60° Geometry: 0 – 100 GU |
Aperture Size | 20° Geometry: 9 × 10 mm; 60° Geometry: 9 × 16 mm |
Repeatability | 20° Geometry: ± 0.2 GU (0–100 GU); 60° Geometry: ± 0.1 GU (0–20 GU), ± 0.2 GU (20–100 GU) |
Reproducibility | 20° Geometry: ± 1.0 GU (0–100 GU); 60° Geometry: ± 0.5 GU (0–20 GU), ± 1.0 GU (20–100 GU) |
Available for Aperture | 20° Geometry: 32 mm, 12 mm; 60° Geometry: 32 mm |